Original Product [Aircraft, General Purpose] by System One Co., Ltd.
Our products are custom-made products complying with specifications of each customer. Our proven technology makes up satisfactory products.
Temperature distribution measuring device
Personal computer
Data logger
Master 3 lines
Heat treatment furnace
Analysis measurement(Max 12lines)
-
[Overview] This device is used for performing verification tests of heat treatment furnaces, combination tests and temperature distribution tests.
Calibration is available according to AMS2750.
-
【Specifications】 Applicable sensor ... R, K, N, J, T
Accuracy: Whichever is larger the whole range ± 0.6 ° C or 0.1% of Reading
(calibration cycle: three months)
Input CH ... Master (for comparison) input 3CH
Control (for control) input 3CH
Distribution input 9 CH (up to 12 CH)
-
[Device configuration] XYZθ stage ... 1 unit
Sensor probe ... Based on flaw detection shape
Sensor amplifier ... Based on the number of sensor probes
Personal computer ... 1 unit
Measurement software ... 1 set (GE approved specification)
Thermocouple inspection device
Personal computer
Measuring instrument
Test thermocouple 6 lines
Temperature furnace
-
[Overview] This device is used for inspection of thermocouples and measuring the correction value of thermocouples for each temperature.
Calibration is available according to AMS2750.
-
[Specifications] Applicable sensor ... R, K, N, J, T
Accuracy: Whichever is larger the whole range ± 0.6 ° C or 0.1% of Reading
Input CH ... 6 CH
-
[Device configuration] Laptop computer ... 1 unit
Data logger ... 1 unit
Measurement software ... 1 set
-
[Software requirements] Windows XP or higher level (Windows 7 compatible)
CPU ... Dual core and higher level recommended
Memory ... 2GB and more recommended
HDD ... 2GB and more free space
*Output test reports are customizable upon customers' requests.
-
[Calibration] Calibration is available according to in AMS2750.NIST traceability compliance
LVDT measuring device
- [Overview] This device is for measuring with high accuracy, moving LVDT by specified length
-
[Specifications] Applicable sensor ... LVDT
Display resolution ... 0.001 mm
Accuracy: ± 0.01 mm (Calibration cycle: 12 months)
-
[Device configuration] Drive unit ... 1 unit
Length indicator ... 1 unit
- [Calibration] NIST traceability